This instruction manual has been written for the use of the Olympus Polarizing Microscope Model BHSP. It is recommended to read the manual carefully in order to familiarize yourself fully with the use of the microscope, so that you can obtain optimum performance from it.
ASSEMBLY
This picture illustrates the sequential procedure of assembly. The numbers indicate the order of assembly of various components. Remove dust caps before mounting components. Take care to keep all glass surfaces clean, and avoid scratching the glass surfaces.
NOTE: For numbers 0, 0, 0 and 0 please refer to explanations in detail on the next page.
Summary of Putting the Microscope in Operation
Model BHSP
Match the line voltage selector switch to local mains voltage (see page 10).
Switch on the light source.
Place a specimen slide on the stage.
Remove the Bertrand lens and analyzer from the light path.
Coarse focus with the 10X objective.
Make interpupillary and diopter adjustments (page 10).
Set the analyzer to optimum extinction position (page 11).
Center the condenser (page 12).
Center the stage (page 13).
J . Center objectives other than 10X (page 13).
K. Swing in the desired objective.
L. Set the condenser, analyzer and Bertrand lens correctly according to your microscopic purpose (pages 15 and 16)
M. Fine focus.
N. Adjust aperture iris diaphragm and field iris diaphragm (page 13).